An Ultrasonically Actuated Silicon-Microprobe-Based Testicular Tubule Assay
- 21 July 2009
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 56 (11), 2666-2674
- https://doi.org/10.1109/tbme.2009.2026910
Abstract
Microdissection testicular sperm extraction (TESE) is an invasive surgical procedure in which sparsely located healthy larger diameter tubules carrying viable spermatazoa are identified by visual examination of the seminiferous tubules of the infertile testis under a microscope, and biopsies of regions of interest are performed. In this paper, we report on microfabricated silicon microprobes integrated with an ultrasonic horn actuator and strain gauges for microdissection probe-TESE (MP-TESE) surgery. The microprobes, with axial-force-sensitive polysilicon strain gauges, have high force sensitivity (-0.4 V/N). The probes were used to detect the boundaries between seminiferous tubules, thus enabling identification of individual tubule diameters. Insertion experiments were performed on rat testis tissue, and by monitoring the tubule puncture in the recorded force, we were able to estimate the average diameter ~ 41.2 plusmn 1.6 mum of the sperm-carrying tubules in samples. We have also demonstrated the ability to sense the existence of larger tubules embedded in a mess of thinner tubules, using an analytically calculated expression for the distribution of sizes measured by the microprobe. This information is important in MP-TESE to distinguish tubules with and without fertile sperm, potentially eliminating the large incision currently required for optical spermatazoa localization.Keywords
This publication has 25 references indexed in Scilit:
- Ultrasonically Actuated Silicon Microprobes for Cardiac Signal RecordingIEEE Transactions on Biomedical Engineering, 2006
- Force Modeling for Needle Insertion Into Soft TissueIEEE Transactions on Biomedical Engineering, 2004
- A multifunctional silicon-based microscale surgical systemSensors and Actuators A: Physical, 2001
- Testicular sperm extraction with intracytoplasmic sperm injection for nonobstructive azoospermiaUrology, 1997
- Plasma-etched neural probesSensors and Actuators A: Physical, 1997
- Silicon-based microelectrodes for neurophysiology fabricated using a gold metallization/nitride passivation systemJournal of Microelectromechanical Systems, 1996
- High fertilization and pregnancy rate after intracytoplasmic sperm injection with spermatozoa obtained from testicle biopsyHuman Reproduction, 1995
- An implantable multielectrode array with on-chip signal processingIEEE Journal of Solid-State Circuits, 1986
- Morphometric studies on rat seminiferous tubulesJournal of Anatomy, 1982