Carbonized blood deposited on fibres during 810, 940 and 1,470 nm endovenous laser ablation: thickness and absorption by optical coherence tomography
Open Access
- 10 February 2010
- journal article
- research article
- Published by Springer Science and Business Media LLC in Lasers in Medical Science
- Vol. 25 (3), 439-447
- https://doi.org/10.1007/s10103-009-0749-1
Abstract
Endovenous laser ablation (EVLA) is commonly used to treat saphenous varicosities. Very high temperatures at the laser fibre tip have been reported during EVLA. We hypothesized that the laser irradiation deposits a layer of strongly absorbing carbonized blood of very high temperature on the fibre tip. We sought to prove the existence of these layers and study their properties by optical transmission, optical coherence tomography (OCT) and microscopy. We analysed 23 EVLA fibres, 8 used at 810 nm, 7 at 940 nm and 8 at 1,470 nm. We measured the transmission of these fibres in two wavelength bands (450–950 nm; 950–1,650 nm). We used 1,310 nm OCT to assess the thickness of the layers and the attenuation as a function of depth to determine the absorption coefficient. Microscopy was used to view the tip surface. All fibres showed a slightly increasing transmission with wavelength in the 450–950 nm band, and a virtually wavelength-independent transmission in the 950–1,650 nm band. OCT scans showed a thin layer deposited on all 13 fibres investigated, 6 used at 810 nm, 4 at 940 nm and 3 at 1,470 nm, some with inhomogeneities over the tip area. The average absorption coefficient of the 13 layers was 72 ± 16 mm−1. The average layer thickness estimated from the transmission and absorption measurements was 8.0 ± 2.7 µm. From the OCT data, the average maximal thickness was 26 ± 6 µm. Microscopy of three fibre tips, one for each EVLA wavelength, showed rough, cracked and sometimes seriously damaged tip surfaces. There was no clear correlation between the properties of the layers and the EVLA parameters such as wavelength, except for a positive correlation between layer thickness and total delivered energy. In conclusion, we found strong evidence that all EVLA procedures in blood filled veins deposit a heavily absorbing hot layer of carbonized blood on the fibre tip, with concomitant tip damage. This major EVLA mechanism is unlikely to have much wavelength dependence at similar delivered energies per centimetre of vein. Optical–thermal interaction between the vein wall and the transmitted laser light depends on wavelength.Keywords
This publication has 17 references indexed in Scilit:
- Intraluminal Fibre-Tip Centring can Improve Endovenous Laser Ablation: A Histological StudyEuropean Journal of Vascular and Endovascular Surgery, 2010
- Is a vein filled with blood a good model for studying endovenous laser ablation?Lasers in Surgery and Medicine, 2009
- Endovenous Laser Ablation–Induced Complications: Review of the Literature and New CasesDermatologic Surgery, 2009
- Heat conduction from the exceedingly hot fiber tip contributes to the endovenous laser ablation of varicose veinsLasers in Medical Science, 2009
- In Vivo Three-Dimensional Imaging of Neovascular Age-Related Macular Degeneration Using Optical Frequency Domain Imaging at 1050 nmInvestigative Ophthalmology & Visual Science, 2008
- Mathematical modeling of 980‐nm and 1320‐nm endovenous laser treatmentLasers in Surgery and Medicine, 2007
- Comparison of Endovenous Radiofrequency Versus 810 nm Diode Laser Occlusion of Large Veins in an Animal ModelDermatologic Surgery, 2002
- Temperature along the surface of modified fiber tips for Nd:YAG laser angioplastyLasers in Surgery and Medicine, 1991
- Refractive index of some mammalian tissues using a fiber optic cladding methodApplied Optics, 1989
- A three-zone model of soft-tissue damage by a CO2laserPhysics in Medicine & Biology, 1986