Sources of inaccuracy in the measurement of adult patients’ resting blood pressure in clinical settings
Top Cited Papers
Open Access
- 1 March 2017
- journal article
- review article
- Published by Ovid Technologies (Wolters Kluwer Health) in Journal of Hypertension
- Vol. 35 (3), 421-441
- https://doi.org/10.1097/hjh.0000000000001197
Abstract
To interpret blood pressure (BP) data appropriately, healthcare providers need to be knowledgeable of the factors that can potentially impact the accuracy of BP measurement and contribute to variability between measurements. A systematic review of studies quantifying BP measurement inaccuracy. Medline and CINAHL databases were searched for empirical articles and systematic reviews published up to June 2015. Empirical articles were included if they reported a study that was relevant to the measurement of adult patients’ resting BP at the upper arm in a clinical setting (e.g. ward or office); identified a specific source of inaccuracy; and quantified its effect. Reference lists and reviews were searched for additional articles. A total of 328 empirical studies were included. They investigated 29 potential sources of inaccuracy, categorized as relating to the patient, device, procedure or observer. Significant directional effects were found for 27; however, for some, the effects were inconsistent in direction. Compared with true resting BP, significant effects of individual sources ranged from −23.6 to +33 mmHg SBP and −14 to +23 mmHg DBP. A single BP value outside the expected range should be interpreted with caution and not taken as a definitive indicator of clinical deterioration. Where a measurement is abnormally high or low, further measurements should be taken and averaged. Wherever possible, BP values should be recorded graphically within ranges. This may reduce the impact of sources of inaccuracy and reduce the scope for misinterpretations based on small, likely erroneous or misleading, changes.Keywords
This publication has 178 references indexed in Scilit:
- Blood Pressure Monitoring Technique Impacts Hypertension TreatmentJournal of General Internal Medicine, 2011
- Comparative Assessment of Four Blood Pressure Measurement Methods in HypertensivesCanadian Journal of Cardiology, 2011
- How Important is the Recommended Slow Cuff Pressure Deflation Rate for Blood Pressure Measurement?Annals of Biomedical Engineering, 2011
- Acute effects of coffee on endothelial function in healthy subjectsEuropean Journal of Clinical Nutrition, 2010
- Evaluating the Accuracy of an Aneroid Sphygmomanometer in a Clinical Trial SettingAmerican Journal of Hypertension, 2009
- White coat effect and white coat hypertension in pediatric patientsPediatric Nephrology, 2002
- Effects of caffeine on pressor regulation during rest and exercise in men at risk for hypertensionAmerican Heart Journal, 1991
- Blood Pressure in the Paretic Arms of Patients with StrokeThe New England Journal of Medicine, 1986
- Acute hemodynamic effects of cigarette smoking in man assessed by systolic time intervals and echocardiography.Circulation, 1979
- Acute effects of low doses of alcohol on left ventricular function by echocardiography.Circulation, 1975