Thin film nanocalorimeter for heat capacity measurements of 30 nm films
- 1 June 2009
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 80 (6), 063901
- https://doi.org/10.1063/1.3142463
Abstract
A silicon nitride membrane-based nanocalorimeter is described for measuring the heat capacity of 30 nm films from 300 mK to 800 K and in high magnetic fields with absolute accuracy ∼2%. The addenda heat capacity of the nanocalorimeter is less than 2×10−7 J/K at room temperature and 2×10−10 J/K at 2.3 K. This is more than ten times smaller than any existing calorimeter suitable for measuring thin films over this wide temperature range. The heat capacities of thin Cu and Au films are reported and agree with bulk values. The thermal conductivity of the thin low stress silicon nitride is substantially smaller than thicker membranes while the specific heat is enhanced below 20 K. Design of the nanocalorimeter will be discussed along with fabrication details and calibration results.Keywords
This publication has 22 references indexed in Scilit:
- Thermodynamic measurements of submilligram bulk samples using a membrane-based “calorimeter on a chip”Review of Scientific Instruments, 2008
- Si-N membrane-based microcalorimetry: Heat capacity and thermal conductivity of thin filmsThermochimica Acta, 2005
- Specific heat and thermal conductivity of low-stress amorphous Si–N membranesSolid State Communications, 2004
- Numerical simulation of the heat transfer in amorphous silicon nitride membrane-based microcalorimetersReview of Scientific Instruments, 2003
- Critical behavior ofPhysical Review B, 2002
- Specific heat of endohedral and higher fullerene thin filmsThe Journal of Chemical Physics, 1999
- Heat capacity measurements of Sn nanostructures using a thin-film differential scanning calorimeter with 0.2 nJ sensitivityApplied Physics Letters, 1997
- Size-Dependent Melting Properties of Small Tin Particles: Nanocalorimetric MeasurementsPhysical Review Letters, 1996
- High-speed (104 °C/s) scanning microcalorimetry with monolayer sensitivity (J/m2)Applied Physics Letters, 1995
- Thin film microcalorimeter for heat capacity measurements from 1.5 to 800 KReview of Scientific Instruments, 1994