Robust water/fat separation in the presence of large field inhomogeneities using a graph cut algorithm
Open Access
- 26 October 2009
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 63 (1), 79-90
- https://doi.org/10.1002/mrm.22177
Abstract
Water/fat separation is a classical problem for in vivo proton MRI. Although many methods have been proposed to address this problem, robust water/fat separation remains a challenge, especially in the presence of large amplitude of static field inhomogeneities. This problem is challenging because of the nonuniqueness of the solution for an isolated voxel. This paper tackles the problem using a statistically motivated formulation that jointly estimates the complete field map and the entire water/fat images. This formulation results in a difficult optimization problem that is solved effectively using a novel graph cut algorithm, based on an iterative process where all voxels are updated simultaneously. The proposed method has good theoretical properties, as well as an efficient implementation. Simulations and in vivo results are shown to highlight the properties of the proposed method and compare it to previous approaches. Twenty‐five cardiac datasets acquired on a short, wide‐bore scanner with different slice orientations were used to test the proposed method, which produced robust water/fat separation for these challenging datasets. This paper also shows example applications of the proposed method, such as the characterization of intramyocardial fat. Magn Reson Med, 2010.Keywords
This publication has 37 references indexed in Scilit:
- Quantification of hepatic steatosis with MRI: The effects of accurate fat spectral modelingJournal of Magnetic Resonance Imaging, 2009
- Multiecho dixon fat and water separation method for detecting fibrofatty infiltration in the myocardiumMagnetic Resonance in Medicine, 2008
- Multiecho water‐fat separation and simultaneous R estimation with multifrequency fat spectrum modelingMagnetic Resonance in Medicine, 2008
- Relaxation effects in the quantification of fat using gradient echo imagingMagnetic Resonance Imaging, 2008
- Joint estimation of water/fat images and field inhomogeneity mapMagnetic Resonance in Medicine, 2008
- Bayesian parallel imaging with edge‐preserving priorsMagnetic Resonance in Medicine, 2006
- What energy functions can be minimized via graph cuts?Ieee Transactions On Pattern Analysis and Machine Intelligence, 2004
- Exact optimization for markov random fields with convex priorsIEEE Transactions on Pattern Analysis and Machine Intelligence, 2003
- Fast approximate energy minimization via graph cutsIeee Transactions On Pattern Analysis and Machine Intelligence, 2001
- Spatial resolution properties of penalized-likelihood image reconstruction: space-invariant tomographsIEEE Transactions on Image Processing, 1996