Stability Analysis of an Enhanced Load Sharing Posterior Fixation Device and Its Equivalent Conventional Device in a Calf Spine Model
- 1 November 1999
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in Spine
- Vol. 24 (21), 2206
- https://doi.org/10.1097/00007632-199911010-00006
Abstract
An in vitro test of calf spine lumbar segments to compare biomechanical stabilization of a rigid versus a dynamic posterior fixation device. To compare flexibility of a dynamic pedicle screw fixation device with an equivalent rigid device. Dynamic pedicle screw device studies are not as prevalent in the literature as studies of rigid devices. These devices contain the potential to enhance load sharing and optimize fusion potential while maintaining stability similar to that of rigid systems. Load-displacement tests were performed on intact and stabilized calf spines for the dynamic and rigid devices. Stability across a destabilized L3–L4 segment was restored by insertion of either a 6 mm × 40 mm dynamic or rigid pedicle screw fixation device across the L2–L4 segment. The screws then were removed, 7 mm × 45 mm pedicle screws of the opposite type were inserted, and the construct then was re-tested. Axial pull-out tests were performed to assess the likely effects of pedicle screw replacement on the load-displacement data. Results indicated a 65% reduction in motion in flexion–extension and a 90% reduction in lateral bending across the destabilized level for both devices, compared with intact spine values. Reduction in axial rotation motion was much smaller than in other modes. Axial pull-out tests showed no weakening of the bone–screw interface. Both devices provided significant stability of similar magnitudes in flexion, extension, and lateral bending. In axial rotation, the devices only could restore stability to levels similar to those in an intact spine. The dynamic device offers a design that may enhance load sharing without sacrificing construct stability.Keywords
This publication has 22 references indexed in Scilit:
- Biomechanical Evaluation of Anterior and Posterior Fixations in an Unstable Calf Spine ModelSpine, 1997
- An In Vitro Study of the Biomechanical Effects of Flexible Stabilization on the Lumbar SpineSpine, 1997
- Biomechanical comparison of calf and human spinesJournal of Orthopaedic Research, 1996
- Static and Cyclical Biomechanical Analysis of Pedicle Screw Spinal ConstructsSpine, 1993
- Longitudinal Element Size Effect on Load Sharing, Internal Loads, and Fatigue Life of Tri-Level Spinal Implant ConstructsSpine, 1993
- Kinematics of the Lumbar Spine Following Pedicle Screw Plate FixationSpine, 1993
- Influence of Bone Mineral Density on the Fixation of Thoracolumbar Implants A Comparative Study of Transpedicular Screws, Laminar Hooks, and Spinous Process WiresSpine, 1990
- An Analytical Investigation of the Mechanics of Spinal InstrumentationSpine, 1988
- A Technique to Evaluate an Internal Spinal Device by Use of the Selspot System: An Application to Luque Closed LoopSpine, 1987