Forehead Pulse Oximetry

Abstract
Background This study investigated whether a tensioning headband that applies up to 20 mmHg pressure over a forehead pulse oximetry sensor could improve arterial hemoglobin oxygen saturation reading accuracy in presence of venous pooling and pulsations at the forehead site. Methods Healthy volunteers were studied breathing room air in supine and various levels of negative incline (Trendelenburg position) using the forehead sensor with the headband adjusted to its maximum and minimum recommended pressure limits. Saturation readings obtained from the forehead sensor with the subjects supine and the headband in place were used as a baseline to compare the effects of negative incline on reading accuracy when using and not using the headband. Occurrences of false low-saturation readings detected by forehead sensors were compared with those from digit sensors. Results No difference was observed between saturation readings obtained from the forehead sensor in supine and negative incline positions when the headband was applied. Forehead sensor readings obtained while subjects were inclined and the headband was not used were significantly lower (P < 0.05) than the supine readings. There was no statistically significant difference between the digit and forehead sensor in reporting false low-saturation readings when the headband was applied, regardless of body incline. Conclusions Application of up to 20 mmHg pressure on the forehead pulse oximetry sensor using an elastic tensioning headband significantly reduced reading errors and provided consistent performance when subjects were placed between supine and up to 15 degrees head-down incline (Trendelenburg position).