Microchip Flow Cytometry Using Electrokinetic Focusing

Abstract
Flow cytometry of fluorescently labeled and unlabeled latex particles is demonstrated on a microfabricated device. The latex particles were detected and counted using laser light scattering and fluorescence coincidence measurements. Sample confinement was accomplished using electrokinetic focusing at a cross intersection, and detection occurred 50 μm downstream from the intersection. Particles with diameters of 1 and 2 μm were analyzed and distinguished from each other based on their light scattering intensity and fluorescence. A maximum sample throughput of 34 particles/s was achieved. Sample mixtures with varying proportions of fluorescently labeled and unlabeled particles were also analyzed and found to be within experimental error of the expected ratios.