A national survey on radiation dose in CT in The Netherlands
Open Access
- 15 May 2013
- journal article
- research article
- Published by Springer Science and Business Media LLC in Insights into Imaging
- Vol. 4 (3), 383-390
- https://doi.org/10.1007/s13244-013-0253-9
Abstract
Objectives To assess radiation exposure due to CT in the Netherlands. Methods Twenty-one hospitals participated in a dose survey for the 21 most frequently used CT protocols. Hospitals completed a Web survey with detailed parameters for one patient per protocol, including the dose length product (DLP) from the scanner dose report. Only standard-sized patients (1.74 m and 77 kg and BMI 25.4 kg/m2 ± 15 %) for each protocol and available scanner were considered. Effective dose (E) per protocol was estimated using ICRP-103-based E/DLP coefficients. Dose levels were compared to surveys from other countries and to diagnostic reference levels. Results Data of 186 patients (247 scan phases) from 14 hospitals and 19 scanners were used for final analysis of DLP and E. Effective doses varied from 0.2 mSv in sinus CT to 19.4 mSv for multiphase liver. The most frequent exams were brain (1.5 mSv), abdomen (8.0 mSv), and thorax-abdomen (11.5 mSv). These results are lower than in Germany and comparable to those in the UK, and are within reference levels. Results between hospitals varied, with per protocol minimum/maximum E ratios ranging from 1.1–5.4. Conclusions Compared to surrounding countries, CT in the Netherlands is associated with relatively low radiation doses in standard patients. Important differences remain between hospitals. Main Messages • A national dose survey providing updated, detailed data for patient dose in the most frequently used CT protocols. • CT in the Netherlands is associated with relatively low individual radiation doses in standard patients compared to surrounding European countries. • Considerable differences remain between hospitals for the most frequently used CT protocols, indicating the need for further optimisation.Keywords
This publication has 10 references indexed in Scilit:
- Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length ProductRadiology, 2010
- Overranging at Multisection CT: An Underestimated Source of Excess Radiation ExposureRadioGraphics, 2010
- Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007Archives of Internal Medicine, 2009
- Radiation Dose Associated With Common Computed Tomography Examinations and the Associated Lifetime Attributable Risk of CancerArchives of Internal Medicine, 2009
- Radiologic and Nuclear Medicine Studies in the United States and Worldwide: Frequency, Radiation Dose, and Comparison with Other Radiation Sources—1950–2007Radiology, 2009
- Overranging in Multisection CT: Quantification and Relative Contribution to Dose—Comparison of Four 16-Section CT ScannersRadiology, 2007
- National survey of doses from CT in the UK: 2003The British Journal of Radiology, 2006
- An inter-hospital comparison of patient dose based on clinical indicationsEuropean Radiology, 2006
- Radiation doses to patients receiving computed tomography examinations in British Columbia.2006