Visibility of microcalcification clusters and masses in breast tomosynthesis image volumes and digital mammography: A 4AFC human observer study
- 12 April 2012
- journal article
- radiation imaging-physics
- Published by Wiley in Medical Physics
- Vol. 39 (5), 2431-2437
- https://doi.org/10.1118/1.3694105
Abstract
Purpose : To investigate the visibility of simulated lesions in digital breast tomosynthesis (BT) image volumes compared with 2D digital mammography (DM). Methods : Simulated lesions (masses and microcalcifications) were added to images of the same women acquired on a DM system (Mammomat Novation, Siemens) and a BT prototype. The same beam quality was used for the DM and BT acquisitions. The total absorbed dose resulting from a 25‐projection BT acquisition and reconstruction (BT25) was approximately twice that of a single DM view. By excluding every other projection image from the reconstruction (BT13), approximately the same dose as in DM was effected. Simulated microcalcifications were digitally added with varying contrast to the DM and BT images. Simulated masses with 8 mm diameter were also added to BT images. A series of 4‐alternative forced choice (4AFC) human observer experiments were conducted. Four medical physicists participated in all experiments, each consisting of 60 trials per experimental condition. The observers interpreted the BT image volumes in cine‐mode at a fixed image sequence speed. The required threshold contrast (St) to achieve a detectability index (d′) of 2.5 (i.e., 92.5% correct decisions) was determined. Results : The St for mass detection in DM was approximately a factor of 2 higher than required in BT indicating that the detection of masses was improved under BT conditions compared to DM. St for microcalcification detection was higher for BT than for DM at both BT dose levels (BT25 and BT13), with a statistically significant difference in St between DM and BT13. These results indicate a dose‐dependent decrease in detection performance in BT for detection of microcalcifications. Conclusions : In agreement with previous investigations, masses of size 8 mm can be detected with less contrast in BT than in DM indicating improved detection performance for BT. However, for the investigated microcalcifications, the results of this study indicate potentially worse performance for BT than for DM at the same dose level.Keywords
This publication has 21 references indexed in Scilit:
- In‐plane visibility of lesions using breast tomosynthesis and digital mammographyMedical Physics, 2010
- Image artifacts in digital breast tomosynthesis: Investigation of the effects of system geometry and reconstruction parameters using a linear system approachMedical Physics, 2008
- A new approach to digital breast tomosynthesis for breast cancer screeningPublished by SPIE-Intl Soc Optical Eng ,2007
- Analyzing the effect of dose reduction on the detection of mammographic lesions using mathematical observer modelsPublished by SPIE-Intl Soc Optical Eng ,2006
- Adaptation of Image Quality Using Various Filter Setups in the Filtered Backprojection Approach for Digital Breast TomosynthesisLecture Notes in Computer Science, 2006
- Digital breast tomosynthesis using an amorphous selenium flat panel detectorPublished by SPIE-Intl Soc Optical Eng ,2005
- Human observer detection experiments with mammograms and power‐law noiseMedical Physics, 2001
- Estimation of the noisy component of anatomical backgroundsMedical Physics, 1999
- Theoretical framework for filtered back projection in tomosynthesisPublished by SPIE-Intl Soc Optical Eng ,1998
- Digital tomosynthesis in breast imaging.Radiology, 1997