The Light–Tissue Interaction of Pulse Oximetry
- 1 December 2007
- journal article
- review article
- Published by Ovid Technologies (Wolters Kluwer Health) in Anesthesia & Analgesia
- Vol. 105 (6), S10-S17
- https://doi.org/10.1213/01.ane.0000269522.84942.54
Abstract
The underlying science of pulse oximetry is based on a simple manipulation of the Lambert-Beer law, which describes the attenuation of light traveling through a mixture of absorbers. Signals from detected red and infrared light that has traveled through blood-perfused tissues are used to estimate the underlying arterial hemoglobin oxygen saturation. However, light scatters in tissue and influences some of the simplifications made in determining this relationship. Under most clinical circumstances, the empirical process that manufacturers use to calibrate the system during its design readily accommodates this and results in accurate readings. The same tissue light scattering properties allow sensors to be configured for use on opposing or adjacent surfaces, provided that the placement sites offer sufficient signal strength and are absent factors known to influence accuracy. In this paper I review the light-tissue interaction in pulse oximetry and describe some of the assumptions made and their implications. Certain deviations from the nominal conditions, whether clinical in nature or misuse of the product, can affect system performance. Consequently, users should be cautious in modifying sensors and/or using them on tissue sites not intended by the manufacturer (off-label use). While perhaps helpful for obtaining pulsatile signals or extending the lifetime of a sensor, some practices can disrupt the optical integrity of the measurement and negatively impact the oxygen saturation reading accuracy.This publication has 31 references indexed in Scilit:
- Forehead Pulse OximetryAnesthesiology, 2006
- Influence of blood vessels on the measurement of hemoglobin oxygenation as determined by time-resolved reflectance spectroscopyMedical Physics, 1995
- Diffusion-based model of pulse oximetry: in vitro and in vivo comparisonsApplied Optics, 1994
- Yet Another Place for the Pulse Oximeter ProbeAnesthesia & Analgesia, 1993
- The theoretical basis for the determination of optical pathlengths in tissue: temporal and frequency analysisPhysics in Medicine & Biology, 1992
- Oral Pulse Oximetry in Small ChildrenAnesthesia & Analgesia, 1991
- Effect of anemia on pulse oximeter accuracy at low saturationJournal of Clinical Monitoring and Computing, 1990
- The Penumbra EffectAnesthesiology, 1989
- Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical propertiesApplied Optics, 1989
- Statistics of Penetration Depth of Photons Re-emitted from Irradiated TissueJournal of Modern Optics, 1989