Biomechanical Characterization of Keratoconus Corneas Ex Vivo With Brillouin Microscopy
Top Cited Papers
Open Access
- 1 July 2014
- journal article
- cornea
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Investigative Ophthalmology & Visual Science
- Vol. 55 (7), 4490-4495
- https://doi.org/10.1167/iovs.14-14450
Abstract
Purpose.: Loss of corneal strength is a central feature of keratoconus progression. However, it is currently difficult to measure corneal mechanical changes noninvasively. The objective of this study is to evaluate if Brillouin optical microscopy can differentiate the mechanical properties of keratoconic corneas versus healthy corneas ex vivo. Methods.: We obtained eight tissue samples from healthy donor corneas used in Descemet's stripping endothelial keratoplasty (DSEK) and 10 advanced keratoconic corneas from patients undergoing deep anterior lamellar keratoplasty (DALK). Within 2 hours after surgery, a confocal Brillouin microscope using a monochromatic laser at 532 nm was used to map the Brillouin frequency shifts of the corneas. Results.: The mean Brillouin shift in the anterior 200 μm of the keratoconic corneas at the cone was measured to be 7.99 ± 0.10 GHz, significantly lower than 8.17 ± 0.06 GHz of the healthy corneas (P < 0.001). The Brillouin shift in the keratoconic corneas decreased with depth from the anterior toward posterior regions with a steeper slope than in the healthy corneas (P < 0.001). Within keratoconic corneas, the Brillouin shift in regions away from the apex of the cone was significantly higher than within the cone region (P < 0.001). Conclusions.: Brillouin measurements revealed notable differences between healthy and keratoconic corneas. Importantly, Brillouin imaging showed that the mechanical loss is primarily concentrated within the area of the keratoconic cone. Outside the cone, the Brillouin shift was comparable with that of healthy corneas. The results demonstrate the potential of Brillouin microscopy for diagnosis and treatment monitoring of keratoconus.This publication has 40 references indexed in Scilit:
- Genome-wide association analyses identify multiple loci associated with central corneal thickness and keratoconusNature Genetics, 2013
- In vivo structural imaging of the cornea by polarization-resolved second harmonic microscopyBiomedical Optics Express, 2011
- Ocular Response Analyzer Measurements in Keratoconus with Normal Central Corneal Thickness Compared with Matched Normal Control EyesJournal of Refractive Surgery, 2011
- Corneal Biomechanical Metrics and Anterior Segment Parameters in Mild KeratoconusOphthalmology, 2010
- Cross-axis cascading of spectral dispersionOptics Letters, 2008
- Confocal Brillouin microscopy for three-dimensional mechanical imagingNature Photonics, 2007
- Assessment of the Biomechanical Properties of the Cornea with the Ocular Response Analyzer in Normal and Keratoconic EyesInvestigative Ophthalmology & Visual Science, 2007
- EMMPRIN and MMP-1 in KeratoconusCornea, 2006
- Determining in vivo biomechanical properties of the cornea with an ocular response analyzerJournal of Cataract & Refractive Surgery, 2005
- Dynamic corneal imagingJournal of Cataract & Refractive Surgery, 2005