Optimization and validation of methods for mapping of the radiofrequency transmit field at 3T
Open Access
- 22 June 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 64 (1), 229-238
- https://doi.org/10.1002/mrm.22421
Abstract
MRI techniques such as quantitative imaging and parallel transmit require precise knowledge of the radio‐frequency transmit field (B). Three published methods were optimized for robust B mapping at 3T in the human brain: three‐dimensional (3D) actual flip angle imaging (AFI), 3D echo‐planar imaging (EPI), and two‐dimensional (2D) stimulated echo acquisition mode (STEAM). We performed a comprehensive comparison of the methods, focusing on artifacts, reproducibility, and accuracy compared to a reference 2D double angle method. For the 3D AFI method, the addition of flow‐compensated gradients for diffusion damping reduced the level of physiological artifacts and improved spoiling of transverse coherences. Correction of susceptibility‐induced artifacts alleviated image distortions and improved the accuracy of the 3D EPI imaging method. For the 2D STEAM method, averaging over multiple acquisitions reduced the impact of physiological noise and a new calibration method enhanced the accuracy of the B maps. After optimization, all methods yielded low noise B maps (below 2 percentage units), of the nominal flip angle value (p.u.) with a systematic bias less than 5 p.u. units. Full brain coverage was obtained in less than 5 min. The 3D AFI method required minimal postprocessing and showed little sensitivity to off‐resonance and physiological effects. The 3D EPI method showed the highest level of reproducibility. The 2D STEAM method was the most time‐efficient technique. Magn Reson Med, 2010.This publication has 37 references indexed in Scilit:
- Rapid radiofrequency field mapping in vivo using single-shot STEAM MRIMagnetic Resonance in Medicine, 2008
- Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurementsMagnetic Resonance in Medicine, 2008
- Reduction of transmitter B1 inhomogeneity with transmit SENSE slice‐select pulsesMagnetic Resonance in Medicine, 2007
- Saturated double‐angle method for rapid B1+ mappingMagnetic Resonance in Medicine, 2006
- Inherent insensitivity to RF inhomogeneity in FLASH imagingMagnetic Resonance in Medicine, 2004
- Image Distortion Correction in fMRI: A Quantitative EvaluationNeuroImage, 2002
- A precise and user-independent quantification technique for regional comparison of single volume proton MR spectroscopy of the human brainNMR in Biomedicine, 2000
- Functional Magnetic Resonance Neuroimaging Data Acquisition TechniquesNeuroImage, 1996
- Radiofrequency map of an NMR coil by imagingMagnetic Resonance Imaging, 1993
- The solution of the bloch equations in the presence of a varying B1 field—An approach to selective pulse analysisJournal of Magnetic Resonance (1969), 1979