Statistical process control for radiotherapy quality assurance
- 22 August 2005
- journal article
- Published by Wiley in Medical Physics
- Vol. 32 (9), 2777-2786
- https://doi.org/10.1118/1.2001209
Abstract
Every quality assurance process uncovers random and systematic errors. These errors typically consist of many small random errors and a very few number of large errors that dominate the result. Quality assurance practices in radiotherapy do not adequately differentiate between these two sources of error. The ability to separate these types of errors would allow the dominant source(s) of error to be efficiently detected and addressed. In this work, statistical process control is applied to quality assurance in radiotherapy for the purpose of setting action thresholds that differentiate between random and systematic errors. The theoretical development and implementation of process behavior charts are described. We report on a pilot project is which these techniques are applied to daily output and flatness/symmetry quality assurance for a 10 MV photon beam in our department. This clinical case was followed over 52 days. As part of our investigation, we found that action thresholds set using process behavior charts were able to identify systematic changes in our daily quality assurance process. This is in contrast to action thresholds set using the standard deviation, which did not identify the same systematic changes in the process. The process behavior thresholds calculated from a subset of the data detected a 2% change in the process whereas with a standard deviation calculation, no change was detected. Medical physicists must make decisions on quality assurance data as it is acquired. Process behavior charts help decide when to take action and when to acquire more data before making a change in the process.Keywords
This publication has 32 references indexed in Scilit:
- Use of EPID for leaf position accuracy QA of dynamic multi‐leaf collimator (DMLC) treatmentMedical Physics, 2004
- A new approach in dose measurement and error analysis for narrow photon beams (beamlets) shaped by different multileaf collimators using a small detectorMedical Physics, 2004
- EDR2 film dosimetry for IMRT verification using low‐energy photon filtersMedical Physics, 2004
- A procedure to determine the radiation isocenter size in a linear acceleratorMedical Physics, 2004
- Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66Medical Physics, 2003
- American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planningMedical Physics, 1998
- AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45Medical Physics, 1994
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994
- Effects of dosimetric and clinical uncertainty on complication-free local tumor controlRadiotherapy and Oncology, 1988
- Dosimetric Precision Requirements in Radiation TherapyActa Radiologica: Oncology, 1984