Thin-Film Microbiosensors for Glucose−Lactate Monitoring

Abstract
A miniaturized device for simultaneous measurement of glucose and lactate levels was produced by means of photopatterning of enzyme-containing photosensitive membrane precursors. This device shows no cross-talk and a lifetime for both the glucose and the lactate sensors of more than 2 weeks when continuously operated in undiluted bovine serum. Linear response ranges of up to 40 mM for glucose and 25 mM for l-lactate, in combination with 95% response times of 0.98 with respect to laboratory techniques. Subcutaneous measurements of glucose levels in pigs were close to the corresponding blood levels obtained without in vivo calibration.