ON THE OPTICAL MEASUREMENT OF CORNEAL THICKNESS
- 1 October 1980
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 58 (5), 760-766
- https://doi.org/10.1111/j.1755-3768.1980.tb06689.x
Abstract
The optical principle for estimating [human] corneal thickness with a commonly employed method is considered. The theoretical relationship between the apparent and true corneal thickness is described in a simple manner. Theoretically, the physiological variations in refractive index and radius of the cornea induce an insignificant error in the corneal thickness estimate. The theoretical relationship between apparent and true corneal thickness was compared to the performance of 2 commercially available pachometers (Haag-Streit and Zeiss). The actual reading of the pachometers was lower than the theoretical reading by a non-linear error amounting to about 0.010 and 0.100 mm at a true corneal thickness of 0.50 and 1.00 mm, respectively.Keywords
This publication has 4 references indexed in Scilit:
- A TECHNICAL IMPROVEMENT OF THE HAAG‐STREIT PACHOMETER Short CommunicationActa Ophthalmologica, 1977
- ON THE OPTICAL MEASUREMENT OF CORNEAL THICKNESSActa Ophthalmologica, 1971
- Measurement of Corneal Thickness With the Haag-Streit PachometerAmerican Journal of Ophthalmology, 1968
- Tiefenmessung der menschlichen Vorderkammer mit planparallelen Platten (Zusatzgerät zur Spaltlampe)Albrecht von Graefes Archiv für Ophthalmologie, 1952