Nanoscale LC-MS(n): technical design and applications to peptide and protein analysis
- 1 June 2002
- journal article
- research article
- Published by Wiley in Journal of Separation Science
- Vol. 25 (9), 557-568
- https://doi.org/10.1002/1615-9314(20020601)25:9<557::aid-jssc557>3.0.co;2-f
Abstract
We describe a nanoscale capillary column switching HPLC–ESI/MS(n) system in detail. It is intended for routine operation with 50 μm ID packed, fused silica columns. The system accepts large injection volumes (typically 10 μL) and uses a trapping column for fast sample loading. Analytes are transferred and then separated on an analytical column at a flow rate of 100–125 nL/min. The design of the system focuses on both robustness for unattended operation and minimization of extracolumn dead volumes. Although the flow direction is controlled by switching valves, the analytes do not pass through any of these valves except the injection valve. A standard, high-pressure-mixing binary pump (not necessarily designed for nanoscale applications) is used, while the columns, splitter restrictors and electrospray emitters are home made. This provides maximum flexibility for a wide array of applications. Examples of protein and peptide identification at a concentration level of 100 amol/μL are given, including on-column derivatization. In addition, a minor modification of the setup provides the permanent option of using peak parking for interactive data acquisition (e. g. MS(n) experiments during analyte elution).Keywords
This publication has 43 references indexed in Scilit:
- PROTEIN STRUCTURE BY MASS SPECChemical & Engineering News, 2000
- Differential Protein Expression in Phenotypic Variants of Streptococcus pneumoniaeInfection and Immunity, 2000
- The Putative Proteinase Maturation Protein A of Streptococcus pneumoniae Is a Conserved Surface Protein with Potential To Elicit Protective Immune ResponsesInfection and Immunity, 2000
- A single naturally processed measles virus peptide fully dominates the HLA-A*0201-associated peptide display and is mutated at its anchor position in persistent viral strainsEuropean Journal of Immunology, 2000
- Contribution of mass spectrometry to contemporary immunologyMass Spectrometry Reviews, 1998
- Durable Gold-Coated Fused Silica Capillaries for Use in Electrospray Mass SpectrometryAnalytical Chemistry, 1995
- An approach to correlate tandem mass spectral data of peptides with amino acid sequences in a protein databaseJournal of the American Society for Mass Spectrometry, 1994
- Electrospray and Taylor-Cone theory, Dole's beam of macromolecules at last?International Journal of Mass Spectrometry and Ion Processes, 1994
- Capillary liquid chromatography/mass spectrometryMass Spectrometry Reviews, 1994
- Characterization of Peptides Bound to the Class I MHC Molecule HLA-A2.1 by Mass SpectrometryScience, 1992