Effect of force on ablation depth for a XeCl excimer laser beam delivered by an optical fiber in contact with arterial tissue under saline
- 1 January 1992
- journal article
- research article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 12 (6), 576-584
- https://doi.org/10.1002/lsm.1900120603
Abstract
The effect of force applied to a 430 üm single fiber, delivering 60 pulses of 308 nm XeC1 laser radiation at 20 Hz, on the ablation depth in porcine aortic tissue under saline has been investigated. Energy densities of 8,15,25, 28, 31, 37, and 45 mJ/mm2 were used. Force was applied by adding weights from 0 to 10 grams to the fiber. The fiber penetration was monitored by means of a position transducer. At 0 grams, the ablation depth increased linearly with incident energy density, but the fiber did not penetrate the tissue; with any weight added, the fiber penetrated the tissue at energy densities above 15 mJ/mm2. The fiber did not penetrate during the first several pulses, possibly due to gas trapped under the fiber. After these first pulses, a smooth linear advancement of the fiber began, which lasted until the pulse train stopped. The ablation depth increased with increasing energy densities and weights. This effect was largest above 25 mJ/mm2 where the ablation efficiencies (unit mm3 /J), with weights added to the fiber, were substantially larger than values found in 308 nm ablation experiments described in the literature, which were conducted with either a focused laser beam or a fiber without additional force. The results imply that in 308 nm excimer laser angioplasty, force must be applied to the beam delivery catheter for efficient recanalization, and that experiments performed with a focused beam or without actual penetration of the fiber do not represent the situation encountered in excimer laser angioplasty.Keywords
This publication has 15 references indexed in Scilit:
- National Heart, Lung, and Blood Institute Percutaneous Transluminal Coronary Angioplasty Registry as a standard for comparison of new devices. When should we use it, and what should we compare?Circulation, 1991
- Percutaneous excimer laser coronary angioplasty of lesions not ideal for balloon angioplasty.Circulation, 1991
- Some Laser‐Tissue Interactions in 308 nm Excimer Laser Coronary AngioplastyJournal of Interventional Cardiology, 1990
- Percutaneous coronary excimer laser angioplasty in patients with stable and unstable angina pectoris. Acute results and incidence of restenosis during 6-month follow-up.Circulation, 1990
- Pulsed laser angioplasty: Wavelength power and energy dependencies relevant to clinical applicationLasers in Surgery and Medicine, 1988
- Tissue interactions and measurement of ablation rates with ultraviolet and visible lasers in canine and human arteriesLasers in Surgery and Medicine, 1987
- The excimer laser: Gross, light microscopic and ultrastructural analysis of potential advantages for use in laser therapy of cardiovascular diseaseJournal of the American College of Cardiology, 1985
- Pulsed ultraviolet lasers and the potential for safe laser angioplastyThe American Journal of Surgery, 1985
- Laser ablation of human atherosclerotic plaque without adjacent tissue injuryJournal of the American College of Cardiology, 1985
- Far‐ultraviolet laser ablation of atherosclerotic lesionsLasers in Surgery and Medicine, 1984