Ferrous sulphate gels for determination of absorbed dose distributions using MRI technique: basic studies

Abstract
Two gels have been found to be suitable to load with ferrous sulphate solution. In these soft tissue equivalent phantoms, the absorbed dose distribution can be measured after irradiation in clinically used MR imaging equipment. The present studies were carried out using a 0.25 T NMR analyser without imaging properties. A ferrous sulphate solution, 0.05 M with respect to sulphuric acid, can be gelled with 4% gelatin to give a dosemeter which has a response which is linearly correlated (r=0.998) with the absorbed dose in the interval 0-40 Gy. Ferrous sulphate solution can also be gelled with 1% agarose, but this gel has to be purged with oxygen to obtain a linear relationship (r=0.997) in the same absorbed dose interval. The ferrous sulphate loaded gels have a sensitivity which is a factor of 2.2 or 4.0 times higher for gelatin and agarose, respectively, than the ordinary dosemeter solution.