Femoral Artery Complications Associated with the Mynx Closure Device
Open Access
- 10 June 2010
- journal article
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 31 (9), 1737-1740
- https://doi.org/10.3174/ajnr.a2153
Abstract
BACKGROUND AND PURPOSE: Devices to close a femoral arteriotomy are frequently used after catheterization for interventional radiology and cardiac procedures to decrease the time to hemostasis and ambulation and, potentially, to decrease local complications. The Mynx vascular closure device uses a sealant designed to occlude the access tract, resulting in hemostasis. MATERIALS AND METHODS: We retrospectively reviewed all cases in which the Mynx device was used and for which follow-up angiography was available. A total of 146 devices were deployed in 135 patients. A follow-up vascular study visualizing the femoral artery was performed in 26 patients (27 studies). RESULTS: There were 5 (5/27, 18%) cases of intravascular Mynx sealant on follow-up vascular imaging. Three pseudoaneurysms (3/27, 11%) were identified. CONCLUSIONS: In this small study, intravascular sealant and pseudoaneurysms were found frequently after femoral arterial closure with the Mynx vascular closure device.Keywords
This publication has 2 references indexed in Scilit:
- Contrast media and clotting: What is the evidence?Catheterization and Cardiovascular Interventions, 2010
- The safety and efficacy of an extravascular, water‐soluble sealant for vascular closure: Initial clinical results for Mynx™Catheterization and Cardiovascular Interventions, 2007