Abstract
Dosimetric detectors used in high-energy photon radiation dosimetry mainly perform a zero- or one-dimensional measurement. These low-dimensional methods are not always adequate in the context of conformal radiotherapy. Therefore, two-dimensional film dosimetry has attracted attention. We studied a 12-bit CCD-based film digitizer (Vidar VXR-12) with regard to accurate film dosimetry. We investigated the stability, linearity, noise, effects of aberrant light scatter and built-in conversion tables. A digitizing resolution of 75 dpi and a digitizing speed of 20 ms/line result in an optimal signal-to-noise ratio. At optical densities above 2.0, the reading accuracy of the digitizer is limited by noise. The results of various experiments prove both the capabilities and limitations of the digitizer studied. We also propose a method to acquire and process film data using such a digitizer.