Reducing Radiation Exposure in Cardiac Catheterizations for Congenital Heart Disease
- 12 December 2018
- journal article
- research article
- Published by Springer Science and Business Media LLC in Pediatric Cardiology
- Vol. 40 (3), 638-649
- https://doi.org/10.1007/s00246-018-2039-9
Abstract
Ionizing radiation exposure is a necessary risk entailed during congenital cardiac catheterizations. The congenital catheterization lab at Yale New Haven Children’s Hospital employed quality improvement strategies to minimize radiation exposure in this vulnerable population. In two phases, we implemented six interventions, which included adding and utilizing lower fluoroscopy and digital angiography (DA) doses, increasing staff and physician radiation awareness, focusing on tighter collimation, and changing the default fluoroscopy and DA doses to lower settings. Post-intervention data were collected prospectively for all procedures in the congenital catheterization lab and compared to pre-intervention radiation data collected retrospectively. Radiation exposure was measured in total air kerma (mGy), dose area product per body weight (DAP/kg) (µGy m2/kg), and fluoroscopy time (min). Data were collected for a total of 312 cases. In considering all procedures, the DAP/kg decreased by 67.6% and air kerma decreased by 63%. Fluoroscopy time did not change over the study period. Significant decreases in radiation exposure (DAP/kg) by procedure type were seen for atrial septal defect, patent ductus arteriosus, and transcatheter pulmonary valve procedures with a 45%, 42% and 83% decrease, respectively. Air kerma decreased significantly for ASD and PDA procedures with an 80% and 72% decrease, respectively. When compared to national benchmarks, the median DAP/kg and air kerma for these procedures are lower at our institution. The decreases continue to be sustained 2 years post-interventions. Systems-based interventions can be readily implemented in the congenital cardiac catheterization lab with dramatic and sustainable radiation dose reduction for patients.Keywords
This publication has 22 references indexed in Scilit:
- An analysis of radiation dose reduction in paediatric interventional cardiology by altering frame rate and use of the anti-scatter gridJournal of Radiological Protection, 2013
- Quantification of temporal, procedural, and hardware‐related factors influencing radiation exposure during pediatric cardiac catheterizationCatheterization and Cardiovascular Interventions, 2012
- Characterization of radiation exposure and effect of a radiation monitoring policy in a large volume pediatric cardiac catheterization labCatheterization and Cardiovascular Interventions, 2011
- Radiation safety program for the cardiac catheterization laboratoryCatheterization and Cardiovascular Interventions, 2011
- Radiation Dose to the Pediatric Cardiac Catheterization and Intervention PatientAmerican Journal of Roentgenology, 2010
- Cumulative patient effective dose and acute radiation-induced chromosomal DNA damage in children with congenital heart diseaseHeart, 2009
- Diagnostic reference levels and effective dose in paediatric cardiac catheterizationThe British Journal of Radiology, 2007
- The ALARA concept in pediatric cardiac catheterization: techniques and tactics for managing radiation dosePediatric Radiology, 2006
- Patient-Specific Dose and Radiation Risk Estimation in Pediatric Cardiac CatheterizationCirculation, 2005
- ACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures: A report of the American College of Cardiology Foundation/American Heart Association/American College of Physicians Task Force on Clinical Competence and TrainingJournal of the American College of Cardiology, 2004