A Monte Carlo and physical phantom evaluation of quantitative In-111 SPECT
- 24 August 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (17), 4169-4185
- https://doi.org/10.1088/0031-9155/50/17/018
Abstract
Accurate estimation of the 3D in vivo activity distribution is important for dose estimation in targeted radionuclide therapy (TRT). Although SPECT can potentially provide such estimates, SPECT without compensation for image degrading factors is not quantitatively accurate. In this work, we evaluated quantitative SPECT (QSPECT) reconstruction methods that include compensation for various physical effects. Experimental projection data were obtained using a GE VH/Hawkeye system and an RSD torso phantom. Known activities of In-111 chloride were placed in the lungs, liver, heart, background and two spherical compartments with inner diameters of 22 mm and 34 mm. The 3D NCAT phantom with organ activities based on clinically derived In-111 ibritumomab tiuxetan data was used for the Monte Carlo (MC) simulation studies. Low-noise projection data were simulated using previously validated MC simulation methods. Fifty sets of noisy projections with realistic count levels were generated. Reconstructions were performed using the OS-EM algorithm with various combinations of attenuation (A), scatter (S), geometric response (G), collimator-detector response (D) and partial volume compensation (PVC). The QSPECT images from the various combinations of compensations were evaluated in terms of the accuracy and precision of the estimates of the total activity in each organ. For experimental data, the errors in organ activities for ADS and PVC compensation were less than 6.5% except the smaller sphere (-11.9%). For the noisy simulated data, the errors in organ activity for ADS compensation were less than 5.5% except the lungs (20.9%) and blood vessels (15.2%). Errors for other combinations of compensations were significantly (A, AS) or somewhat (AGS) larger. With added PVC, the error in the organ activities improved slightly except for the lungs (11.5%) and blood vessels (3.6%) where the improvement was more substantial. The standard deviation/mean ratios were all less than 1.5%. We conclude that QSPECT methods with appropriate compensations provided accurate In-111 organ activity estimates. For the collimator used, AGS was almost as good as ADS and may be preferable due to the reduced reconstruction time. PVC was important for small structures such as tumours or for organs in close proximity to regions with high activity. The improved quantitative accuracy from QSPECT methods has the potential for improving organ dose estimations in TRT.Keywords
This publication has 30 references indexed in Scilit:
- A new method for modeling the spatially-variant, object-dependent scatter response function in SPECTPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Efficient fully 3-D iterative SPECT reconstruction with Monte Carlo-based scatter compensationIEEE Transactions on Medical Imaging, 2002
- Combination of MCNP and SimSET for Monte Carlo simulation of SPECT with medium- and high-energy photonsIEEE Transactions on Nuclear Science, 2002
- Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECTIEEE Transactions on Medical Imaging, 2002
- Monte Carlo evaluation of object shape effects in iodine-131 SPET tumor activity quantificationEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- SPECT scatter modelling in non-uniform attenuating objectsPhysics in Medicine & Biology, 1997
- A practical method for incorporating scatter in a projector-backprojector for accurate scatter compensation in SPECTIEEE Transactions on Nuclear Science, 1993
- A fast projector-backprojector pair modeling the asymmetric, spatially varying scatter response function for scatter compensation in SPECT imagingIEEE Transactions on Nuclear Science, 1993
- Object shape dependent PSF model for SPECT imagingIEEE Transactions on Nuclear Science, 1993
- Quantitative SPECT imaging with indium-111IEEE Transactions on Nuclear Science, 1991