Image Formation in the Scanning Microscope

Abstract
Fourier imaging in the scanning microscope is considered. It is shown that there are two geometries of the microscope, which have been designated Type 1 and Type 2. Those of Type 1 exhibit identical imaging to the conventional microscope, whereas those of Type 2 (confocal microscopes) display various differences. Imaging of a single point object, two-point resolution and response to a straight edge are also considered. The effect of various arrangements using lenses with annular pupil functions is also discussed. It is found that Type 2 microscopes have improved imaging properties over conventional microscopes and that these may be further improved by use of one or two lenses with annular pupils.