Measuring Computed Tomography Scanner Variability of Radiomics Features
Top Cited Papers
- 1 November 2015
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in Investigative Radiology
- Vol. 50 (11), 757-765
- https://doi.org/10.1097/rli.0000000000000180
Abstract
Objectives The purpose of this study was to determine the significance of interscanner variability in CT image radiomics studies. Materials and Methods We compared the radiomics features calculated for non–small cell lung cancer (NSCLC) tumors from 20 patients with those calculated for 17 scans of a specially designed radiomics phantom. The phantom comprised 10 cartridges, each filled with different materials to produce a wide range of radiomics feature values. The scans were acquired using General Electric, Philips, Siemens, and Toshiba scanners from 4 medical centers using their routine thoracic imaging protocol. The radiomics feature studied included the mean and standard deviations of the CT numbers as well as textures derived from the neighborhood gray-tone difference matrix. To quantify the significance of the interscanner variability, we introduced the metric feature noise. To look for patterns in the scans, we performed hierarchical clustering for each cartridge. Results The mean CT numbers for the 17 CT scans of the phantom cartridges spanned from −864 to 652 Hounsfield units compared with a span of −186 to 35 Hounsfield units for the CT scans of the NSCLC tumors, showing that the phantom's dynamic range includes that of the tumors. The interscanner variability of the feature values depended on both the cartridge material and the feature, and the variability was large relative to the interpatient variability in the NSCLC tumors for some features. The feature interscanner noise was greatest for busyness and least for texture strength. Hierarchical clustering produced different clusters of the phantom scans for each cartridge, although there was some consistent clustering by scanner manufacturer. Conclusions The variability in the values of radiomics features calculated on CT images from different CT scanners can be comparable to the variability in these features found in CT images of NSCLC tumors. These interscanner differences should be considered, and their effects should be minimized in future radiomics studies.Keywords
This publication has 27 references indexed in Scilit:
- Radiomics: Extracting more information from medical images using advanced feature analysisEuropean Journal of Cancer, 2012
- Intratumor Heterogeneity Characterized by Textural Features on Baseline 18F-FDG PET Images Predicts Response to Concomitant Radiochemotherapy in Esophageal CancerJournal of Nuclear Medicine, 2011
- The biology underlying molecular imaging in oncology: from genome to anatome and back againClinical Radiology, 2010
- Relationship between sharpness and noise in CT images reconstructed with different kernelsRadiation Protection Dosimetry, 2010
- Texture analysis in non-contrast enhanced CT: Impact of malignancy on texture in apparently disease-free areas of the liverEuropean Journal of Radiology, 2009
- Hepatic entropy and uniformity: additional parameters that can potentially increase the effectiveness of contrast enhancement during abdominal CTClinical Radiology, 2007
- Decoding global gene expression programs in liver cancer by noninvasive imagingNature Biotechnology, 2007
- Texture-based classification of atherosclerotic carotid plaquesIEEE Transactions on Medical Imaging, 2003
- Computerized analysis of lesions in US images of the breastAcademic Radiology, 1999
- Texture analysis using gray level run lengthsComputer Graphics and Image Processing, 1975