Factors That Affect Radiofrequency Heat Lesion Size
Open Access
- 1 December 2014
- journal article
- Published by Oxford University Press (OUP) in Pain Medicine
- Vol. 15 (12), 2020-2036
- https://doi.org/10.1111/pme.12566
Abstract
This study aims to compare radiofrequency (RF) heat lesion size across electrodes and generator settings available for interventional pain management. Monopolar lesions are generated ex vivo in animal tissue using sharp cannulae with tip diameters 23, 22, 20, 18, 16 gauge; tip lengths 5, 6, 10, 15 mm; set temperatures 60, 70, 80, 90°C; set times 1, 1.5, 2, 3, 5, 10 minutes. Lesions are generated using the RRE electrode, cooled RF, and parallel-tip bipolar RF for comparison. Lesion sizes are assessed by automated photographic temperature inference from over 400 lesions, using multiple lesions per configuration. Monopolar lesion width and length increase with each factor (P < 0.001). Increasing cannula diameter from 22 to 16 gauge increases average lesion width 58–65% (3–4 mm) at 80°C and 2 minutes. Increasing temperature from 60°C to 90°C increases lesion width 108–152% at 2 minutes. Although dimensions grow most rapidly over the first minute, average lesion width is 11–20% larger at 2 minutes, and 23–32% larger at 3 minutes, compared with 1 minute. Lesion length extends distal and proximal to the tip, and exceeds tip length by 1–5 mm at 80°C and 2 minutes. Conventional 16 gauge cannulae at 80–90°C for 2–3 minutes generate lesions of average width similar to that produced by the cooled RF configuration proposed for sacroiliac joint denervation. Bipolar RF between parallel cannulae produces a rounded brick-shaped lesion of comparable shape to three sequential monopolar lesions generated using the same cannulae and generator settings. Tip gauge, tip length, temperature, and time substantially affect RF lesion size.Keywords
This publication has 35 references indexed in Scilit:
- Textbook of Stereotactic and Functional NeurosurgeryPublished by Springer Science and Business Media LLC ,2009
- The Histologic Effects of Pulsed and Continuous Radiofrequency Lesions at 42°C to Rat Dorsal Root Ganglion and Sciatic NerveSpine, 2005
- Efficacy and Validity of Radiofrequency Neurotomy for Chronic Lumbar Zygapophysial Joint PainSpine, 2000
- Nonselective Nerve Fibre Damage in Peripheral Nerves after Experimental ThermocoagulationActa Neurochirurgica, 1998
- Large-volume tissue ablation with radio frequency by using a clustered, internally cooled electrode technique: laboratory and clinical experience in liver metastases.Radiology, 1998
- Hepatic metastases: percutaneous radio-frequency ablation with cooled-tip electrodes.Radiology, 1997
- Radiofrequency tissue ablation: Increased lesion diameter with a perfusion electrodeAcademic Radiology, 1996
- Radiofrequency neurolysis in a clinical modelJournal of Neurosurgery, 1981
- Size, Form and Structural Peculiarities of Experimental Brain Lesions Obtained by Thermocontrolled RadiofrequencyStereotactic and Functional Neurosurgery, 1965
- Reversible Heat Lesions with Radiofrequency CurrentJournal of Neurosurgery, 1964