Optical artefact characterization and correction in volumetric scintillation dosimetry
- 10 December 2013
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (1), 23-42
- https://doi.org/10.1088/0031-9155/59/1/23
Abstract
The goals of this study were (1) to characterize the optical artefacts affecting measurement accuracy in a volumetric liquid scintillation detector, and (2) to develop methods to correct for these artefacts. The optical artefacts addressed were photon scattering, refraction, camera perspective, vignetting, lens distortion, the lens point spread function, stray radiation, and noise in the camera. These artefacts were evaluated by theoretical and experimental means, and specific correction strategies were developed for each artefact. The effectiveness of the correction methods was evaluated by comparing raw and corrected images of the scintillation light from proton pencil beams against validated Monte Carlo calculations. Blurring due to the lens and refraction at the scintillator tank-air interface were found to have the largest effect on the measured light distribution, and lens aberrations and vignetting were important primarily at the image edges. Photon scatter in the scintillator was not found to be a significant source of artefacts. The correction methods effectively mitigated the artefacts, increasing the average gamma analysis pass rate from 66% to 98% for gamma criteria of 2% dose difference and 2 mm distance to agreement. We conclude that optical artefacts cause clinically meaningful errors in the measured light distribution, and we have demonstrated effective strategies for correcting these optical artefacts.Keywords
This publication has 16 references indexed in Scilit:
- Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector systemMedical Physics, 2012
- Vignette and Exposure Calibration and CompensationIeee Transactions On Pattern Analysis and Machine Intelligence, 2010
- Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, HoustonMedical Physics, 2009
- Exploration of the potential of liquid scintillators for real‐time 3D dosimetry of intensity modulated proton beamsMedical Physics, 2009
- Development of an easy-to-handle range measurement tool using a plastic scintillator for proton beam therapyPhysics in Medicine & Biology, 2006
- The three-dimensional scintillation dosimetry method: test for a106Ru eye plaque applicatorPhysics in Medicine & Biology, 2005
- Radiometric framework for image mosaickingJournal of the Optical Society of America A, 2005
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- Acceleration of iterative image restoration algorithmsApplied Optics, 1997
- Using vanishing points for camera calibrationInternational Journal of Computer Vision, 1990