Effect of finish line location and saliva contamination on the accuracy of crown finish line scanning

Abstract
PurposeIntraoral scanners are used widely as an alternative to conventional impressions, but studies on the influence of finish line location and saliva contamination on scanning trueness are lacking. The purpose of this in vitro study was to evaluate the influence of finish line location and saliva contamination on the scanning trueness of crown finish lines. Materials and methodsThree ivorine teeth were prepared for all-ceramic crowns with finish lines placed equigingivally, 0.5 mm subgingivally, and 1.0 mm subgingivally. A single-cord technique was used for gingival retraction, and a total of 180 intraoral scans were made using two intraoral scanners (Emerald; Planmeca USA Inc., Hoffman Estates, IL, USA & Trios 3; 3Shape A/S, Copenhagen, Denmark). The prepared teeth were separated from the dentoform and scanned using the same intraoral scanners to create reference scans. All scans were imported to the design software (Dental System 2019; 3Shape A/S, Copenhagen, Denmark). After marking the finish lines of prepared teeth, intraoral scans were aligned to the reference scans for comparisons. Vertical and horizontal marginal discrepancies were measured at four different measuring points (buccal, lingual, mesial, and distal) and analyzed. Two-way ANOVA and Tukey HSD tests were used for statistical analysis (alpha = 0.05). ResultsThe average vertical and horizontal discrepancies from various groups ranged from -33 to 440 mu m. For both intraoral scanners, subgingival finish line groups showed greater vertical and horizontal discrepancies compared with equigingival finish line groups. Saliva contamination significantly increased both vertical and horizontal discrepancies for all finish line locations. The discrepancy increases due to saliva contamination were greater for the subgingival groups. ConclusionsSubgingival finish lines were not accurately captured using the intraoral scanners. The presence of saliva significantly reduced scanning trueness, and this was amplified when the finish lines were located subgingivally.