Comparison between the HemoCue® and an automated counter for measuring hemoglobin
Open Access
- 1 August 2004
- journal article
- research article
- Published by FapUNIFESP (SciELO) in Revista de Saúde Pública
- Vol. 38 (4), 585-587
- https://doi.org/10.1590/s0034-89102004000400017
Abstract
To determine the precision and agreement of the hemoglobin (Hb) measurements in capillary and venous blood samples by the HemoCue® and an automated counter. Hb was determined by both equipaments in blood samples of 29 pregnant women. The HemoCue® showed low repeatability of Hb measurements in duplicate in capillary (CR=0.53 g/dL, CV=13.6%) and venous blood (CR=0.53 g/dL, CV=13.6%). Hb measurements in capillary blood were higher than those in venous blood (12.4 and 11.7 g/dL, respectively; pKeywords