Comparison ofkQfactors measured with a water calorimeter in flattening filter free (FFF) and conventional flattening filter (cFF) photon beams
Open Access
- 25 January 2018
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 63 (4), 045023
- https://doi.org/10.1088/1361-6560/aaaa93
Abstract
Recently flattening filter free (FFF) beams became available for application in modern radiotherapy. There are several advantages of FFF beams over conventional flattening filtered (cFF) beams, however differences in beam spectra at the point of interest in a phantom potentially affect the ion chamber response. Beams are also non-uniform over the length of a typical reference ion chamber and recombination is usually larger. Despite several studies describing FFF beam characteristics, only a limited number of studies investigated their effect on kQ factors. Some of those studies predicted significant discrepancies in kQ factors (0.4 % up to 1.0 %) if TPR20,10 based codes of practice (CoPs) would be used. This study addresses the question to which extent kQ factors, based on a TPR20,10 CoP, can be applied in clinical reference dosimetry. It is the first study that compares kQ factors measured directly with an absorbed dose to water primary standard in FFF-cFF pairs of clinical photon beams. This was done with a transportable water calorimeter described elsewhere. The measurements corrected for recombination and beam radial non-uniformity were performed in FFF-cFF beam pairs at 6 MV and 10 MV of an Elekta Versa HD for a selection of three different Farmer-type ion chambers (8 serial numbers). The ratio of measured kQ factors of the FFF-cFF beam pairs were compared with the TPR20,10 CoPs of the NCS and IAEA and the %dd(10)x CoP of the AAPM. For the TPR20,10 based CoPs differences less than 0.23 % were found in kQ factors between the corresponding FFF-cFF beams with standard uncertainties smaller than 0.35 %, while for the %dd(10)x these differences were smaller than 0.46 % and within the expanded uncertainty of the measurements. Based on the measurements made with the equipment described in this study the authors conclude that the kQ factors provided by the NCS-18 and IAEA TRS-398 codes of practice can be applied for flattening filter free beams without additional correction. However, existing codes of practice cannot be applied ignoring the significant volume averaging effect of the FFF beams over the ion chamber cavity. For this a corresponding volume averaging correction must be applied.Keywords
Funding Information
- EURAMET and the European Union (EMPIR 16NRM03 RTNORM)
This publication has 35 references indexed in Scilit:
- IPEM topical report 1: guidance on implementing flattening filter free (FFF) radiotherapyPhysics in Medicine & Biology, 2016
- A water calorimeter for on-site absorbed dose to water calibrations in60Co and MV-photon beams including MRI incorporated treatment equipmentPhysics in Medicine & Biology, 2016
- High spatial resolution dosimetric response maps for radiotherapy ionization chambers measured using kilovoltage synchrotron radiationPhysics in Medicine & Biology, 2015
- Combining tissue‐phantom ratios to provide a beam‐quality specifier for flattening filter free photon beamsMedical Physics, 2014
- Determination of the recombination correction factorkSfor some specific plane-parallel and cylindrical ionization chambers in pulsed photon and electron beamsPhysics in Medicine & Biology, 2006
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Absorbed dose beam quality factors for the dosimetry of high-energy photon beamsPhysics in Medicine & Biology, 1992
- Convective velocity effects on a thermistor in waterJournal of Research of the National Bureau of Standards, 1988
- An Absorbed Dose Water Calorimeter: Theory, Design, and PerformanceJournal of Research of the National Bureau of Standards, 1982
- Electron and Photon Beams from a 50 MeV Racetrack MicrotronActa Radiologica: Oncology, 1980