Abstract
Background— Electromyography-guided phrenic nerve (PN) monitoring using a catheter positioned in a hepatic vein can aid in preventing phrenic nerve palsy (PNP) during cryoballoon ablation for atrial fibrillation. We wanted to evaluate the feasibility and efficacy of PN monitoring during procedures using second-generation cryoballoons. Methods and Results— This study included 140 patients (43 women) in whom pulmonary vein isolation was performed using a second-generation cryoballoon. Electromyography-guided PN monitoring was performed by pacing the right PN at 60 per minute and recording diaphragmatic compound motor action potential (CMAP) via a quadripolar catheter positioned in a hepatic vein. If a 30% decrease in CMAP amplitude was observed, cryoapplication was discontinued with forced deflation to avoid a PNP. Monitoring was unfeasible in 8 of 140 patients (5.7%), PNP occurred in 1. Stable CMAP amplitudes were achieved before ablation in 132 of 140 patients (94.3%). In 18 of 132 patients (13.6%), a 30% decrease in CMAP amplitude occurred and cryoablation was discontinued. Each time, recovery of CMAP amplitude took <60 s. In 9 of 18 cases, a second cryoapplication in the same pulmonary vein was safely performed. We observed no PNP or complication related to electromyography-guided PN monitoring. Conclusions— Electromyography-guided PN monitoring using a catheter positioned in a hepatic vein seems feasible and effective to prevent PNP during cryoballoon ablation using second-generation cryoballoon.Keywords
This publication has 17 references indexed in Scilit:
- Clinical Experience With a Novel Electromyographic Approach to Preventing Phrenic Nerve Injury During Cryoballoon Ablation in Atrial FibrillationCirculation: Arrhythmia and Electrophysiology, 2014
- Recordings of diaphragmatic electromyograms during cryoballoon ablation for atrial fibrillation accurately predict phrenic nerve injuryHeart Rhythm, 2014
- Novel Electromyographic Monitoring Technique for Prevention of Right Phrenic Nerve Palsy During Cryoballoon AblationCirculation: Arrhythmia and Electrophysiology, 2013
- Improved In Vivo Performance of Second‐Generation Cryoballoon for Pulmonary Vein IsolationJournal of Cardiovascular Electrophysiology, 2013
- Cryoballoon Ablation of Pulmonary Veins for Paroxysmal Atrial FibrillationJournal of the American College of Cardiology, 2013
- Long-Term Outcomes After Cryoballoon Pulmonary Vein IsolationJournal of the American College of Cardiology, 2013
- Efficacy and safety of cryoballoon ablation for atrial fibrillation: A systematic review of published studiesHeart Rhythm, 2011
- Phrenic nerve monitoring with diaphragmatic electromyography during cryoballoon ablation for atrial fibrillation: The first human applicationHeart Rhythm, 2011
- Diaphragmatic electromyography during cryoballoon ablation: a novel concept in the prevention of phrenic nerve palsyHeart Rhythm, 2011
- Efficacy and Safety of Circumferential Pulmonary Vein Isolation Using a Novel Cryothermal Balloon Ablation SystemJournal of the American College of Cardiology, 2005