Applications of stimulated echo correction to multicomponent T2 analysis
Open Access
- 19 October 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 67 (6), 1803-1814
- https://doi.org/10.1002/mrm.23157
Abstract
We propose a multicomponent fitting algorithm for multiecho T2 data which allows for correction of T2 distributions in the presence of stimulated echoes. Tracking the population of spins in many coherence pathways via the iterated method of the Extended Phase Graph algorithm allows for accurate quantification of echo magnitudes. The resulting decay curves allow for correction of errors due to nonideal refocusing pulses as a result of inhomogeneities in the B1 transmit field. Non-Negative Least Squares fitting is used to quantify the magnitude of T2 components at various T2 values. This method, allowing calculation of the T2 distribution with simultaneous extraction of the refocusing pulse flip angle, requires no change to image acquisition procedures and no extra data input. Validation by means of both simulations and in vivo data shows excellent interscan reproducibility while vastly improving the accuracy of extracted T2 parameters in voxels where poor B1 homogeneity leads to refocusing pulse flip angles significantly less than 180°. Most notably, myelin water fraction values in these regions are found to have increased consistency and accuracy. Magn Reson Med, 2011.Keywords
This publication has 26 references indexed in Scilit:
- Time-efficient fast spin echo imaging at 4.7 T with low refocusing anglesMagnetic Resonance in Medicine, 2009
- Myelin water imaging: Implementation and development at 3.0T and comparison to 1.5T measurementsMagnetic Resonance in Medicine, 2009
- Myelin water imaging in multiple sclerosis: quantitative correlations with histopathologyMultiple Sclerosis Journal, 2006
- Measurement and correction of transmitter and receiver induced nonuniformities in vivoMagnetic Resonance in Medicine, 2005
- Analysis of wave behavior in lossy dielectric samples at high fieldMagnetic Resonance in Medicine, 2002
- Different magnetization transfer effects exhibited by the short and longT2 components in human brainMagnetic Resonance in Medicine, 2000
- In vivo visualization of myelin water in brain by magnetic resonanceMagnetic Resonance in Medicine, 1994
- Practical T2 quantitation for clinical applicationsJournal of Magnetic Resonance Imaging, 1992
- Quantitative interpretation of NMR relaxation dataJournal of Magnetic Resonance (1969), 1989
- Multiecho imaging sequences with low refocusing flip anglesJournal of Magnetic Resonance (1969), 1988