Characterization of the phantom material Virtual Water™ in high‐energy photon and electron beams
- 14 March 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (4), 876-887
- https://doi.org/10.1118/1.2174186
Abstract
The material Virtual Water has been characterized in photon and electron beams. Range-scaling factors and fluence correction factors were obtained, the latter with an uncertainty of around 0.2%. This level of uncertainty means that it may be possible to perform dosimetry in a solid phantom with an accuracy approaching that of measurements in water. Two formulations of Virtual Water were investigated with nominally the same elemental composition but differing densities. For photon beams neither formulation showed exact water equivalence-the water/Virtual Water dose ratio varied with the depth of measurement with a difference of over 1% at 10 cm depth. However, by using a density (range) scaling factor very good agreement (<0.2%) between water and Virtual Water at all depths was obtained. In the case of electron beams a range-scaling factor was also required to match the shapes of the depth dose curves in water and Virtual Water. However, there remained a difference in the measured fluence in the two phantoms after this scaling factor had been applied. For measurements around the peak of the depth-dose curve and the reference depth this difference showed some small energy dependence but was in the range 0.1%-0.4%. Perturbation measurements have indicated that small slabs of material upstream of a detector have a small (<0.1% effect) on the chamber reading but material behind the detector can have a larger effect. This has consequences for the design of experiments and in the comparison of measurements and Monte Carlo-derived values.Keywords
This publication has 27 references indexed in Scilit:
- Absorbed dose to water reference dosimetry using solid phantoms in the context of absorbed-dose protocolsMedical Physics, 2005
- An experimental and Monte Carlo investigation of the energy dependence of alanine/EPR dosimetry: II. Clinical electron beamsPhysics in Medicine & Biology, 2005
- Use of a range scaling method to determine alanine/water stopping power ratiosRadiation Physics and Chemistry, 2003
- The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibrationPhysics in Medicine & Biology, 2003
- Calculations of electron fluence correction factors using the Monte Carlo code PENELOPEPhysics in Medicine & Biology, 2003
- The effect of waterproofing sleeves on ionization chamber responsePhysics in Medicine & Biology, 1992
- Implication of electron backscattering for electron dosimetryPhysics in Medicine & Biology, 1991
- Dose errors due to charge storage in electron irradiated plastic phantomsMedical Physics, 1984
- Epoxy resin based tissue substitutesThe British Journal of Radiology, 1977