Thirty years of experience with epicardial pacing in children

Abstract
Due to underlying cardiovascular anatomy and size, epicardial pacing may be the preferred method of pacing in small children. To assess long-term safety, we reviewed all epicardial pacemakers implanted in children between 1971 and 2001. We found that 122 patients, with a median age of 5.4 years, had a total of 181 pacemakers and 260 electrodes implanted over a total follow-up of 789 patient-years. Of the total, 12 patients died after the first implantation, with one death attributable to dysfunction of the pacemaker. Reintervention was required in 75 patients after 5.0 ± 3.2 years, due to depletion of the battery in 45 patients (60%), fracture or dysfunction of electrodes in 27 patients (36%), and infection in 3 patients (4%). In univariate analyses, risk factors for reintervention were an approach via a median sternotomy, with a relative risk of 2.3 (p = 0.0087), and an indication for pacing other than atrioventricular block, with a relative risk of 1.7 (p = 0.0314). In multivariate analyses, the approach via the median sternotomy independently predicted the need for reintervention, with a relative risk of 2.1, and 95% confidence intervals from 1.1 to 4.1 (p = 0.0256). The longevity of the second pacemaker and/or its electrode, assessed in 26 patients, was 3.7 ± 2.6 years, not shorter than the first implantation (p = 0.4037). We conclude that epicardial pacing is a reliable means of achieving permanent pacing in children, with low morbidity and mortality. A substantial proportion, nonetheless, requires reintervention within five years, warranting meticulous follow-up.

This publication has 1 reference indexed in Scilit: