Increased Sensitivity to Low Concentration Gadolinium Contrast by Optimized Heavily T2-weighted 3D-FLAIR to Visualize Endolymphatic Space

Abstract
Purpose: To increase the sensitivity of 3-dimensional fluid-attenuated inversion recovery (3D-FLAIR) to low concentration gadolinium (Gd)-based contrast medium, we optimized sequence parameters on a phantom and evaluated the optimized sequence in patients suspicious for endolymphatic hydrops. Materials and Methods: All scans were performed on a 3-tesla magnetic resonance (MR) unit using a 32-channel head coil. We optimized sequence parameters using a phantom filled with diluted Gd and compared the optimized protocol with 3D-FLAIR using conventional turbo spin echo sequence (3D-FLAIR-CONV). Nine patients underwent scanning using the newly optimized sequence and 3D-FLAIR-CONV 4 hours after double-dose administration of intravenous Gd. We subjectively scored separation of endo- and perilymph space and measured contrast-to-noise ratio (CNR) between endo- and perilymph. Results: The optimized sequence in the phantom study consisted of: repetition time, 9000 ms; echo time, 540 ms; inversion time, 2400 ms; low constant readout flip angle, 120° in the later part of the echo train. Image contrast became heavily T2-weighted (hT2W-3D-FLAIR). In patients, we recognized endolymphatic space for both the cochlea and vesti- bule significantly better by hT2W-3D-FLAIR than 3D-FLAIR-CONV (PP2W-3D-FLAIR was more sensitive than the previous method to low concentration of Gd. Visualization of the endolymphatic space by double-dose administration of intravenous Gd would be more reliable using hT2W-3D-FLAIR.