Electromagnetic Navigation in Minimally Invasive Spine Surgery: Results of a Cadaveric Study to Evaluate Percutaneous Pedicle Screw Insertion
Open Access
- 1 January 2008
- journal article
- Published by International Journal of Spine Surgery in The International Journal of Spine Surgery
- Vol. 2 (1), 43-47
- https://doi.org/10.1016/SASJ-2007-0105-RR
Abstract
This cadaveric study compared efficacy and safety of an electromagnetic (EM) guidance system versus conventional fluoroscopy for percutaneous pedicle screw fixation. As percutaneous pedicle screw fixation becomes increasingly common in spinal surgery, intraoperative imaging systems that maximize efficiency while minimizing radiation exposure and inaccurate trajectories will be progressively more important. Published studies have validated the safety of percutaneous screw fixation using conventional fluoroscopic guidance and frameless optical stereotaxy, though EM guidance systems have not been evaluated for percutaneous placement in the lumbosacral spine. The aim of the study was to evaluate the clinical applicability of an EM system for minimally invasive spine fusion in the lumbosacral spine. Five human cadaveric specimens underwent bilateral lumbosacral percutaneous screw fixation from L1 to S1 using conventional anteroposterior (AP) and lateral fluoroscopic techniques on one side and 2-dimesional (2D) EM guidance on each matching side. Intraoperative efficiency was evaluated, and pedicle, vertebral, and critical breach rates were assessed on postoperative computed tomography (CT). Overall mean fluoroscopy time per screw was 58.9 ± 44.7 seconds for conventional fluoroscopy compared to 27.4 ± 13.5 seconds for electromagnetic guidance (P = .0003). Pedicle, vertebral, and critical breach rates for the L1-S1 were 32.1%, 10.7%, and 25.0% for conventional fluoroscopy and 42.8%, 10.7%, and 14.1% for electromagnetic guidance (difference not statistically significant [ns]). In comparing critical breaches in the lumbar spine (L1-L5), there was a significant difference between 2-D EM guidance (0) and CF guidance (6) (P = .02). Two-dimensional EM navigation provides a modality for lumbosacral percutaneous pedicle screw fixation that is more efficient and safer than conventional fluoroscopy. This data provides a foundation for further clinical trials of this technology. Level 5 – Bench ResearchKeywords
This publication has 8 references indexed in Scilit:
- Accuracy of Computerized Frameless Stereotactic Image-Guided Pedicle Screw Placement into Previously Fused Lumbar SpinesSpine, 2005
- Electromagnetic Field-Based Image-Guided Spine Surgery Part One: Results of a Cadaveric Study Evaluating Lumbar Pedicle Screw PlacementSpine, 2003
- Electromagnetic Field-Based Image-Guided Spine Surgery Part Two: Results of a Cadaveric Study Evaluating Thoracic Pedicle Screw PlacementSpine, 2003
- Computer-Assisted Thoracic Pedicle Screw PlacementSpine, 2001
- Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patientsEuropean Spine Journal, 2000
- Image-guided surgery: applications to the cervical and thoracic spine and a review of the first 120 proceduresJournal of Neurosurgery: Spine, 2000
- Clinical Evaluation and Computed Tomography Scan Analysis of Screw Tracts After Percutaneous Insertion of Pedicle Screws in the Lumbar SpineSpine, 2000