Coincidence correction in red blood cell counting
- 1 September 1989
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 34 (9), 1239-1246
- https://doi.org/10.1088/0031-9155/34/9/009
Abstract
A method has been developed for accurate correction of red blood cell count for coincidence in aperture-impedance electronic blood cell counters. It is based on extrapolation of the slope of regression of the counts which are obtained with sequential dilutions of the samples.Keywords