The Repeatability of Corneal Thickness Measures
- 1 November 2000
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in Cornea
- Vol. 19 (6), 792-795
- https://doi.org/10.1097/00003226-200011000-00007
Abstract
To compare the repeatability of three measures of corneal thickness: Orbscan Slitscan pachymetry, ultrasound pachymetry, and optical pachymetry. Twenty normal subjects were tested on three occasions. Two occurred on the same day and the third was on a different day at approximately the same time of day as one of the first two visits. Central corneal thickness of the right eye was measured with a Haag-Streit optical pachymeter, a Humphrey Model 855 ultrasound pachymeter, and the Orbscan system. Day-to-day and same-day repeatability was assessed by calculating the difference between the values from two visits and determining the mean difference, the SD, and the 95% limits of agreement (LoA) (LoA = mean +/- 1.96 SD). Mean (+/- SD) central corneal thickness as measured by each instrument was as follows: 539 +/- 33 microm for optical pachymetry, 542 +/- 33 microm for ultrasound pachymetry, and 596 +/- 40 microm for Orbscan pachymetry. For day-to-day comparisons, optical pachymetry showed the poorest repeatability with 95% LoA of -61 to +32 microm. Ultrasound pachymetry showed better repeatability with 95% LoA of -22 to +24 microm. The Orbscan showed the best repeatability centrally with 95% LoA of -10 to +17 microm. Peripheral Orbscan pachymetry was less repeatable than that measured centrally but still more repeatable than central optical pachymetry. Similar results were found with same-day comparisons. The Orbscan system is the most repeatable technique for measuring corneal thickness but shows a significant bias toward greater corneal thickness measures than both ultrasound and optical pachymetry.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography systemBritish Journal of Ophthalmology, 1999
- Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thicknessOphthalmology, 1999
- Corneal Response to OrthokeratologyOptometry and Vision Science, 1998
- Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetryJournal of Cataract & Refractive Surgery, 1997
- Extended Wear in PerspectiveOptometry and Vision Science, 1997
- The effect of age on human corneal thicknessActa Ophthalmologica, 1993
- Variability of Ultrasonic PachometryOptometry and Vision Science, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Optical Pachometry MeasurementsOptometry and Vision Science, 1984
- PachometryOptometry and Vision Science, 1983