Prospectively Randomized Evaluation of Stented Xenograft Hemodynamic Function in the Aortic Position

Abstract
Background— Standard stented aortic xenograft valves have not yet been compared regarding their hemodynamic function using a stratified intraoperative randomization protocol. Methods and Results— 100 patients were prospectively included after intraoperative metric sizing of the decalcified aortic annulus. They received Mosaic (M) or Perimount (P) aortic valve replacement. Patient age was 73±5 years, 51 were female, and New York Heart Association (NYHA) functional class was 2.8±0.5. The 21-mm annulus group consisted of 5 (M)/7 (P) patients, the 23-mm annulus group of 20 (M)/20 (P), the 25-mm annulus group of 18 (M)/19 (P), and the 27-mm annulus group of 4 (M)/7 (P) patients, respectively. Hemodynamic function was evaluated using transthoracic echocardiography before discharge and at follow-up (438±352 days). Surgery was uncomplicated in all patients. Labeled valve sizes were 0.93 (M) and 1.05 (P) mm smaller than the annulus diameters (p=NS). In-hospital mortality was 5%, all nonvalve-related. Transvalvular...