The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness
- 28 September 2009
- journal article
- research article
- Published by Wiley in Clinical & Experimental Ophthalmology
- Vol. 37 (7), 687-691
- https://doi.org/10.1111/j.1442-9071.2009.02109.x
Abstract
Purpose: To assess agreement between the iCare rebound tonometer and Goldmann tonometry and to assess the influence of central corneal thickness and the value of scleral rebound tonometer readings. Methods: Prospective single‐centre cross‐sectional study comparing iCare rebound tonometer (RT) intraocular pressure (IOP) readings taken from corneal and scleral locations to Goldmann applanation tonometer (GAT) readings in 100 subjects attending ophthalmology clinics. Results: There was a significant difference between RT and GAT, with RT tending to overestimate IOP. The mean difference between RT and GAT measurements was 3.36 mmHg. The mean difference between the log of RT and the log of GAT measurements was 0.2356, a ratio of 1.27 (P < 0.0001). A formula derived from a linear regression analysis suggested that a 10% increase in CCT increased the RT IOP reading by 9.9%. Scleral RT readings showed no relationship to GAT readings. Conclusions: The rebound tonometer cannot replace the Goldmann tonometer in the office setting given the wide limits of agreement between the two devices. Corneal rebound tonometer readings are influenced by CCT whereas scleral rebound tonometer readings are of no value.Keywords
This publication has 34 references indexed in Scilit:
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thicknessBritish Journal of Ophthalmology, 2006
- Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma PatientsJournal of Glaucoma, 2006
- Accuracy of the New ICare Rebound Tonometer vs. Other Portable Tonometers in Healthy EyesOptometry and Vision Science, 2006
- Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal populationOphthalmic and Physiological Optics, 2005
- Three portable tonometers, the TGDc‐01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*Ophthalmic and Physiological Optics, 2005
- The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathologyBritish Journal of Ophthalmology, 2004
- Pressure-volume relation for the living human eyeCurrent Eye Research, 2000
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- AUTOMATIC DETERMINATION OF THE RESOLVING POWER OF THE EYEActa Ophthalmologica, 1971