Abstract
The Leeds physiological anatomical (PA) hip joint simulator was developed to apply three axes of loading and a complex three-dimensional motion so that the forces and motions can reproduce exactly the walking cycles defined by Paul. This paper presents the results of a study using the Leeds PA hip joint simulator to determine the wear of 32 mm ultra-high molecular weight polyethylene (UHMWPE) acetabular cups against stainless steel and zirconia ceramic heads, using bovine serum as lubricant. These results have been compared with the results of a previous study that used water as the lubricant, which led to UHMWPE transfer film being formed on the stainless steel head. Comparisons are also made with clinical results and results from other simulators. The study indicates that it is preferable to use bovine serum in simulator studies. In addition, the results indicate that if the surface roughness of the metallic and femoral heads are similar, and they remain undamaged during the tests, the wear rates of the UHMWPE cups are likely to be similar.