Comparison of Fixation Disparity Measurements Obtained with the Wesson Fixation Disparity Card and the Sheedy Disparometer

Abstract
We compared the fixation disparity measurements obtained with the Disparometer to those of the Wesson Card. Previous studies suggest that the measurements obtained by these two instruments gave different results. The results of previous studies were based on the analysis of pooled data, which may have led to spurious results. Analysis of our data done in this manner (pooled data) revealed the same findings. However, after analysis of the data based on the separation of subjects according to phoria (exophoria/esophoria), no statistically significant difference was found between the measurements obtained with the Wesson Card and those with the Disparometer.