A Novel, Self-Expanding, Nitinol Stent in Medically Refractory Intracranial Atherosclerotic Stenoses

Abstract
Background and Purpose— The purpose of this study was to assess the safety and performance of the Wingspan stent system and Gateway percutaneous transluminal angioplasty balloon catheter in the treatment of high-grade, intracranial atherosclerotic lesions in patients who had failed medical therapy. Methods— In this prospective, multicenter, single-arm study, medically refractory patients with a modified Rankin score ≤3 and recurrent symptoms attributable to angiographically demonstrated intracranial stenosis ≥50% in a vessel 2.5 to 4.5 mm in diameter were enrolled. Intracranial lesions were predilated with an undersized Gateway balloon catheter to 80% of the native vessel diameter, followed by deployment of the self-expanding Wingspan stent to facilitate further remodeling of the atherosclerotic plaque and to maintain vessel patency. Neurologic examinations and angiograms were performed at 6 months after the procedure. Results— Among the 45 patients enrolled, the degree of stenosis was reduced from a base...