Zika virus RNA polymerase chain reaction on the utility channel of a commercial nucleic acid testing system

Abstract
BACKGROUNDSeveral countries have implemented safety strategies to reduce the risk of Zika virus (ZIKV) transmission through blood transfusion. These strategies have included nucleic acid amplification testing (NAT) of blood donations. In this study, a new real-time polymerase chain reaction (PCR) assay including internal control for the detection of ZIKV on the cobas omni Utility Channel (UC) on the cobas 6800 system is presented. STUDY DESIGN AND METHODSPCR conditions and primer/probe concentrations were optimized on the LightCycler 480 instrument. Optimized conditions were transferred to the cobas omni UC on the cobas 6800 system. Subsequently, the limit of detection (LOD) in plasma and urine, genotype inclusivity, specificity, cross-reactivity, and clinical sensitivity were determined. RESULTSThe 95% LOD of the ZIKV PCR assay on the cobas 6800 system was 23.0 IU/mL (95% confidence interval [CI], 16.5-37.5) in plasma and 24.5 IU/mL (95% CI, 13.4-92.9) in urine. The assay detected African and Asian lineages of ZIKV. The specificity was 100%. The clinical concordance between the newly developed ZIKV PCR assay and the investigational Roche cobas Zika NAT test was 83% (24/29). CONCLUSIONSWe developed a sensitive ZIKV PCR assay on the cobas omni UC on the cobas 6800 system. The assay can be used for large-scale screening of blood donations for ZIKV or for testing of blood donors returning from areas with ZIKV to avoid temporal deferral. This study also demonstrates that the cobas omni UC on the cobas 6800 system can be used for in-house-developed PCR assays.