Sirolimus-Eluting Stent for the Treatment of In-Stent Restenosis

Abstract
Background— We have previously reported the safety and effectiveness of sirolimus-eluting stents for the treatment of de novo coronary lesions. The present investigation explored the potential of this technology to treat in-stent restenosis. Methods and Results— Twenty-five patients with in-stent restenosis were successfully treated with the implantation of 1 or 2 sirolimus-eluting Bx VELOCITY stents in São Paulo, Brazil. Nine patients received 2 stents (1.4 stents per lesion). Angiographic and volumetric intravascular ultrasound (IVUS) images were obtained after the procedure and at 4 and 12 months. All vessels were patent at the time of 12-month angiography. Angiographic late loss averaged 0.07±0.2 mm in-stent and −0.05±0.3 mm in-lesion at 4 months, and 0.36±0.46 mm in-stent and 0.16±0.42 mm in-lesion after 12 months. No patient had in-stent or stent margin restenosis at 4 months, and only one patient developed in-stent restenosis at 1-year follow-up. Intimal hyperplasia by 3-dimensional IVUS was 0.92±1.9 mm 3 at 4 months and 2.55±4.9 mm 3 after 1 year. Percent volume obstruction was 0.81±1.7% and 1.76±3.4% at the 4- and 12-month follow-up, respectively. There was no evidence of stent malapposition either acutely or in the follow-up IVUS images, and there were no deaths, stent thromboses, or repeat revascularizations. Conclusion— This study demonstrates the safety and the potential utility of sirolimus-eluting Bx VELOCITY stents for the treatment of in-stent restenosis.