Torque During Canal Instrumentation Using Rotary Nickel-Titanium Files

Abstract
Nickel-titanium engine-driven rotary instruments are used increasingly in endodontic practice. One frequently mentioned problem is fracture of an instrument in the root canal. Very few studies have been conducted on torsional characteristics of these instruments, and none has been done under dynamic conditions. The purposes of this study were to measure the torque generated and the apical force applied during instrumentation with a commercial engine-driven nickel-titanium file system, and to relate torque generated during simulated clinical use to torsional failure of the instruments. Ten extracted human teeth (five with small-sized and five with medium-sized straight root canals) were instrumented with Quantec Series 2000 files, and the torque and apical force generated were measured. The applied apical force was generally low, not exceeding 150 g in either small or medium canals. The torque depended on the tip size and taper of each instrument, and on canal size. Instruments with 0.05 and 0.06 taper generated the highest torque, which was greater in small than in medium canals. The torque at failure was significantly (p < 0.001) higher than torque during instrumentation, but with considerable variation in the extent of the difference.