Hand-held dynamometry in patients with haematological malignancies: Measurement error in the clinical assessment of knee extension strength
Open Access
- 9 March 2009
- journal article
- Published by Springer Science and Business Media LLC in BMC Musculoskeletal Disorders
- Vol. 10 (1), 31
- https://doi.org/10.1186/1471-2474-10-31
Abstract
Hand-held dynamometry is a portable and inexpensive method to quantify muscle strength. To determine if muscle strength has changed, an examiner must know what part of the difference between a patient's pre-treatment and post-treatment measurements is attributable to real change, and what part is due to measurement error. This study aimed to determine the relative and absolute reliability of intra and inter-observer strength measurements with a hand-held dynamometer (HHD). Two observers performed maximum voluntary peak torque measurements (MVPT) for isometric knee extension in 24 patients with haematological malignancies. For each patient, the measurements were carried out on the same day. The main outcome measures were the intraclass correlation coefficient (ICC +/- 95%CI), the standard error of measurement (SEM), the smallest detectable difference (SDD), the relative values as % of the grand mean of the SEM and SDD, and the limits of agreement for the intra- and inter-observer '3 repetition average' and the 'highest value of 3 MVPT' knee extension strength measures. The intra-observer ICCs were 0.94 for the average of 3 MVPT (95%CI: 0.86-0.97) and 0.86 for the highest value of 3 MVPT (95%CI: 0.71-0.94). The ICCs for the inter-observer measurements were 0.89 for the average of 3 MVPT (95%CI: 0.75-0.95) and 0.77 for the highest value of 3 MVPT (95%CI: 0.54-0.90). The SEMs for the intra-observer measurements were 6.22 Nm (3.98% of the grand mean (GM) and 9.83 Nm (5.88% of GM). For the inter-observer measurements, the SEMs were 9.65 Nm (6.65% of GM) and 11.41 Nm (6.73% of GM). The SDDs for the generated parameters varied from 17.23 Nm (11.04% of GM) to 27.26 Nm (17.09% of GM) for intra-observer measurements, and 26.76 Nm (16.77% of GM) to 31.62 Nm (18.66% of GM) for inter-observer measurements, with similar results for the limits of agreement. The results indicate that there is acceptable relative reliability for evaluating knee strength with a HHD, while the measurement error observed was modest. The HHD may be useful in detecting changes in knee extension strength at the individual patient level.Keywords
This publication has 66 references indexed in Scilit:
- Physical exercise as adjuvant therapy for patients undergoing hematopoietic stem cell transplantationBone Marrow Transplantation, 2007
- Enhancing the examiner’s resisting force improves the reliability of manual muscle strength measurements: Comparison of a new device with hand-held dynamometryJournal of Rehabilitation Medicine, 2007
- Evaluation of anaemia in patients with multiple myeloma and lymphoma: findings of the European CANCER ANAEMIA SURVEYEuropean Journal of Haematology, 2006
- What change in isokinetic knee muscle strength can be detected in men and women with hemiparesis after stroke?Clinical Rehabilitation, 2005
- Hand-Held Dynamometry in Persons with TetraplegiaAmerican Journal of Physical Medicine & Rehabilitation, 2005
- Role of a mixed type, moderate intensity exercise programme after peripheral blood stem cell transplantationBritish Journal of Sports Medicine, 2004
- Two strategies of transferring from sit-to-stand; The activation of monoarticular and biarticular musclesJournal of Biomechanics, 1994
- Sources of variation in back and leg dynamometryErgonomics, 1994
- Measuring change over time: Assessing the usefulness of evaluative instrumentsJournal of Chronic Diseases, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986