Evaluation of the performance of elastomeric pumps in practice: are we under-delivering on chemotherapy treatments?
Open Access
- 30 October 2017
- journal article
- research article
- Published by Taylor & Francis Ltd in Current Medical Research and Opinion
- Vol. 33 (12), 2153-2159
- https://doi.org/10.1080/03007995.2017.1374936
Abstract
Background and aims: Elastomeric pumps are widely used to facilitate ambulatory chemotherapy, and studies have shown that they are safe and well received by patients. Despite these advantages, their end of infusion time can fluctuate significantly. The aim of this research was to observe the performance of these pumps in real practice and to evaluate patients’ satisfaction. Methods: This was a two-phase study conducted at three cancer units over 6 months. Phase-1 was an observational study recording the status of pumps at the scheduled disconnection time and noting remaining volume of infusion. Phase-2 was a survey of patients and their perception/satisfaction. Ethical approval was granted. Results: A total of 92 cases were observed covering 50 cases disconnected at hospital and 42 disconnected at home. The infusion in 40% of hospital disconnection cases was slow, with patients arriving at hospital with unfinished pumps; 58% of these had an estimated remaining volume which exceeded 10 mL with 35% exceeded 20 mL. In 73% of these cases, and regardless of the remaining volume, the patient was disconnected and the pump was discarded. Conclusions: The performance of pumps varied, which affected nurse workload and patients’ waiting-times. A smart system is an option to monitor the performance of pumps and to predict their accuracy.Keywords
This publication has 20 references indexed in Scilit:
- Effect of environmental conditions on performance of elastomeric pumps.American Journal of Health-System Pharmacy, 2013
- Improving Wait Times and Patient Satisfaction in Primary CareJournal for Healthcare Quality, 2013
- Effects of atmospheric pressure conditions on flow rate of an elastomeric infusion pumpAmerican Journal of Health-System Pharmacy, 2012
- Evaluation of the design and reliability of three elastomeric and one mechanical infusersJournal of Oncology Pharmacy Practice, 2007
- Willing to wait?: The influence of patient wait time on satisfaction with primary careBMC Health Services Research, 2007
- A phase II study of FOLFIRI-3 (double infusion of irinotecan combined with LV5FU) after FOLFOX in advanced colorectal cancer patientsBritish Journal of Cancer, 2006
- Timeliness in Ambulatory Care TreatmentThe Journal of Ambulatory Care Management, 2003
- The Flow Rate of the Elastomeric Balloon Infusor is Influenced by the Internal Pressure of the InfusorJournal of Korean Medical Science, 2001
- Home infusion therapySeminars in Oncology Nursing, 1996
- An evaluation of six infusion devices for the continuous infusion of cytotoxic drugs in ambulatory patientsJournal of Oncology Pharmacy Practice, 1995