GeneXpert Enterovirus Assay: One-Year Experience in a Routine Laboratory Setting and Evaluation on Three Proficiency Panels

Abstract
A prospective unblinded comparative evaluation of three assays for the detection of enteroviral RNA performed on 83 positive and 79 negative cerebrospinal fluid samples showed initial and resolved sensitivities of 90.4% and 98.8%, respectively, for the Cepheid GeneXpert enterovirus assay; 94.0% and 97.6%, respectively, for the Argene enterovirus consensus kit; and 100% and 100%, respectively, for an in-house real-time PCR. The initial and resolved specificities were 100% for all three assays.