Clinical and Dosimetric Implications of Air Gaps between Bolus and Skin Surface during Radiation Therapy
Open Access
- 1 January 2013
- journal article
- Published by Scientific Research Publishing, Inc. in Journal of Cancer Therapy
- Vol. 04 (07), 1251-1255
- https://doi.org/10.4236/jct.2013.47147
Abstract
Purpose: The main objective of the study was to evaluate the effect of air gaps of 0 - 5.0 cm between bolus and skin for 1.0 cm Superflab bolus on surface dose (DSurf) and depth of maximum dose (dmax) in solid water and Rando® phantoms. Methods: In this work, the effects of bolus to surface distance on DSurf and variation in dmax were analyzed in a solid water phantom and in an anthropomorphic Rando® phantom for different field sizes, using Gafchromic® EBT films and farmer chamber. Results: For field sizes of 5 × 5 cm2 the DSurf is significantly affected by increasing air gaps greater than 5 mm. For field sizes larger than 10 × 10 cm2, DSurf is nearly the same for air gaps of 0 - 5.0 cm. For small fields and 6 MV photon beam, dmax increases with increasing air gap, while for 10 MV beam and smaller field sizes (i.e. 5 × 5 and 10 × 10 cm2) the dmax first decreases and then increases with the air gaps. For both 3DCRT and IMRT plans on Rando®, DSurf reduction is more prominent with increasing air gaps. Conclusion: For field sizes larger than 10 × 10 cm2 DSurf is largely unaffected by air gaps. However, smaller air gap results in shallower dmax for both 6 MV and 10 MV photon beams at all fields sizes. Special consideration should be taken to reduce air gaps between bolus and skin for field sizes smaller than 10 × 10 cm2 or when surface contour variations are greater or when the bolus covers small area and at the border of the field. </pKeywords
This publication has 12 references indexed in Scilit:
- Dose distribution homogeneity in two TBI techniques—Analysis of 208 irradiated patients conducted in Stanislaw Leszczynski Memorial Hospital, KatowiceReports of Practical Oncology and Radiotherapy, 2012
- The influence of the bolus-surface distance on the dose distribution in the build-up regionReports of Practical Oncology and Radiotherapy, 2010
- The accuracy of the pencil beam convolution and anisotropic analytical algorithms in predicting the dose effects due to attenuation from immobilization devices and large air gapsMedical Physics, 2009
- Assessment of skin dose for breast chest wall radiotherapy as a function of bolus materialPhysics in Medicine & Biology, 2008
- Evaluation of surface and superficial dose for head and neck treatments using conventional or intensity-modulated techniquesPhysics in Medicine & Biology, 2007
- Surface buildup dose dependence on photon field delivery technique for IMRTJournal of Applied Clinical Medical Physics, 2004
- Surface and build‐up region dosimetry for obliquely incident intensity modulated radiotherapy 6 MV x raysMedical Physics, 2003
- Doses near the surface during total-body irradiation with 15 MV X-raysInternational Journal of Cancer, 2001
- Beam spoilers versus bolus for 6 mv photon treatment of head and neck cancersMedical Dosimetry, 2000
- Effects on skin dose from unwanted air gaps under bolus in photon beam radiotherapyRadiation Measurements, 2000